Ultraviolet Fluorescence Sulphur Analyzer

Overview
HK-DP-T3000 Ultraviolet fluorescence sulphur analyser adopts advanced total sulfur analysis method-ultra violet fluorescence method, determine the total sulphur content of various kinds of gas, liquid, solid oils and chemical products. It is a apparatus with high sensitivity conforms to ASTM D5453, mainly used for the determination of sulfur in liquid hydrocarbons, boiling in the range from approximately 25 to 400°C, with viscosities between approximately 0.2 and 20 cSt (m m3/s) at room temperature.
Features
· We adopt the most advanced ultraviolet lamp and other critical components to ensure the super high sensitivity and reliability.
· Adopts the imported American Nafion pipe membrane dryer, good water removal to ensure the stability apparatus.
· The apparatus can establish and stroe the working curve, user can make the specimen analysis without establishing the working curve again.
· By selecting the gas sample injector or the solid sample injector, The apparatus can make the analysis of gas, liquid and solid sample to meet various testers,
Technical Parameters
· Specimen species : liquid, solid and gas specimen
· Measurement method : ultraviolet fluorescence
· Specimen sample size : solid sample size : 1~20mg; liquid sample size : 1~30 µL; gas sample size : 1~10 ML
· Measurement range : 0.2 mg/L~3%
· Detection lower limit : 0.2 mg/L
· Temperature control range : room temperature to 1100°C
· Temperature control precision : 0.5%±2°C
· Gas source requirements : argon : above 99.995% oxygen : above 99.99% humidity <5ppm
· Rated voltage : AC220V±22V 50Hz±0.5Hz
· Power : 1500W