Oxygen & Water Vapor Dual Mode Permeation Tester

Introduction
This tester can be
used to do high-precision oxygen permeability and water vapor permeability testing
of film or sheet at the same time;
It adopts
Coulometry theory for oxygen permeability test, and adopt international
advanced electrolytic process theory for water vapor permeability test;
It is applicable
for testing the oxygen transmission property and water vapor transmission
property of plastic film, sheet and plastic bottles and bags etc. packaging containers
in air or pure oxygen;.
Execution technical
standard for oxygen permeability: GB/T 19789-2005, ASTM D 3985-1995, ASTM F
1927-1998, ASTM F 1307-1990, YBB00082003, ISO 15105-2:2003, DIN 53380-3-1998,
JIS K-7126 ;
Execution technical
standard for water vapor permeability: ASTM F1770, ASTM D1653-03, ASTM
E96-00e1, ISO15106-3, JIS Z0208-1976, DIN 53122-1, ISO2528.
Testing
principle
It adopts
Coulometry theory;
The testing
chambers including upper and nether parts, the sample film is fixed between the
two chambers, the high pure oxygen (O2) circulate in the upper chamber of the
film, and the high pure nitrogen (N2) circulate in the nether chamber of film,
the oxygen molecules get through the film and spray to nitrogen in the nether
chamber, and brought by the blowing nitrogen to the sensor, so, the tester then
analyze the oxygen concentration as what the sensor tested and calculate the
oxygen transmission value. But for the packing containers, the nitrogen circulate
in the bottle and the air or oxygen circulate outside the container For water
vapor permeability testing;
It is made
according to the electrolytic process theory;
The testing
chambers including upper and nether parts, the sample film is fixed between the
two chambers, water vapor circulate in the upper chamber of the film, and dry
gas circulate in the nether chamber of film, water molecules get through the
film and spray to dry gas in the nether chamber, and brought by the blowing gas
to the sensor, so, the tester then analyze the water vapor concentration as what
the sensor tested and calculate the water vapor transmission value.