Hydrocarbon, Interfacial Tension & Foaming Tester

GD-11132 LIQUID PETROLEUM
PRODUCTS HYDROCARBON TESTER
Purpose
The
instrument is designed and made as per GB/T11132-2002. It uses fluorescence
indicator to let the hydrocarbon in the liquid petroleum products shown on the
absorption column, and calculates its percentage by volume. It is taken as an
important quality index of engine fuels and aviation fuels.
Main
Parameters
Voltage
adjustment range of decompression valve: 0~400kPa;
Electric
vibrator: controlled by two independent ways;
Ultraviolet
light tube: 1220 mm in length; wave length 365 ± 5nm;
Illumination
light: 1220 mm in length; power is 40 W;
Gas
supply mode; Nitrogen in steel cylinder (or compressed air in steel cylinder);
INTERFACIAL TENSION TESTER
GD-6541A AUTOMATIC INTERFACIAL
TENSION TESTER
Purpose
The
instrument is designed and made as per GB/T6541 “Test Methods for Interfacial
Tension of Petroleum Products between Oils and Water (Ring Methods)â€. It is
suitable to determine interfacial tension between mineral oils and water under
non-equilibrium conditions (Liquid-Liquid interface). It can also be used to
determine surface tension of various liquids (liquid-gas-interface).
Characteristics
Adopts
a large screen dot matrix LCD.
Automatic
temperature compensation, clock control, automatic storage for suddenly power
off, print, and average value calculation.
Technical
Parameters
Measurement
range: (1) For insulating oils: 2 ~ 100mN/m;(2) For other liquids: 2 ~ 200mN/m;
Sensitive:
0.1mN/m; Accuracy:0.1mN/m; Resolution: 0.1mN/n; Repeatability: 0.3 %
FOAMING TESTER
GD-12579 Lubricating Oils Foaming
Characteristics Tester
Purpose
The
Instrument is designed and made as per GB/T12579, ASTM D 892 and AP 146, It is
suitable to determine foaming characteristics of lubricating oils, so that to
evaluate the foaming tendency and foaming stability of lubricating oils.
Technical
Specification
Main
heating power: 65 W; Auxiliary heating power: 1000W; Refrigeration power: 500
W;
Air
flow rate: 94 ± 5ml/min, adjustable; Temperature controlling range for high
temperature bath;
Room
temperature to 99.9°C: Temperature controlling range
For
low temperature bath: 5 to 99.9°C; Temperature controlling accuracy: ±0.5°C;