Gas Chromatograph hal 4

- Types of injectors : packed column injection, capillary tube injection, six-way valve
gas injection, automatic sampler ;
- Number of detectors : 3 (maximum); FID , TCD , ECD , FPD and NPD optional ;
- Start sample injection : manual and automatic are optional;
- Communication interface : Ethernet: IEEE802.3 ;
B. Detector Specifications
- Hydrogen flame ionization detector (FID)
- Linear range : ≥106
- Detection limit : ≤5×10-12 g/s(hexadecane/isooctane)
- Baseline noise : ≤ 0.05 P A
- Baseline drift : ≤1x10-13A/30min
- Hydrogen Generator:
1. Mains supply voltage : 220v 50Hz
2. Maximum power : 150W; 180W
3. Ambient temperature : 0-40?, relative humidity <85%.
4. Environmental conditions : without much dust and corrosive gas.
5. Gas purity : 99.999% (relative oxygen content)
6. Output flow : Type 300 0-300 m1/min; Type 500 0-500 m1/min
7. Output pressure : 0.4Mpa (if special pressure is required, notification should be
given tofactory before leaving the factory to determine other things)
- Thermal Conductivity Detector (TCD)
- Baseline noise : ≤ 20 μV
- Linear range : ≥104
- Baseline drift : ≤ 3 0μ V /30min
- Sensitivity : S≥ 10000 mV•ml/mg ( benzene/toluene ) ( With magnification of 1, 2,
3, 4 times optional )